Turning on the System
1) Fire the arc lamp under the microscope table
(make sure nothing else is on)
2) Turn on the brightfield bulb power
switch on the microscope base (#2)
3) Turn on the camera next to the
monitor(#3)
4) Log on and start the MicroManager software. Use the “. . .” button to
change the configuration the PVCAM simple configuration.
5) Write down your start time on the log sheet
Shutting down the System
The arc lamp must be on for at least 20-30 min prior to turning it off, you must wait that long before
turning it back on, too. Firing the arc greatly shortens the lamp life, so it’s best to leave it running during
the day when the schedule shows other users are coming—check the schedule, the last person MUST shut
down the system.
If there is another user after you:
1) Clean any oil or water immersion objectives you used during your session.
2) Log off from your Windows session
3) Write down the time on the log sheet.
If you are the last person scheduled:
1) Clean any oil or water immersion objectives you used during your session.
2) Log off your Windows session, but leave the computer running (don’t shutdown).
3) Turn off the camera (#3), brightfield power (#2), and the arc lamp (#1).
4) Cover the scope with the plastic cover. THE LAMP HOUSING MAY BE HOT! Don’t cover that part.
5) Write down the time on the log sheet (how else will the next person know if system is cool
enough to restart?)