LSM 510 Meta Training Notes
Single Track
z Select Channel Mode for normal
scanning and Single Track. Follow the
excitation/emission pathway in the
image to define your configuration.
z Press Excitation to select laser line
and % transmission.
z Press and select an Excitation
dichroic; this selection depends on the
laser lines being used to excite the
sample.
z Select 1
0
and 2
0
dichroics; selection is
based upon wavelengths you want to
capture. Dichroics reflect short
wavelengths and pass long wavelengths.
z Activate detectors; ChS, Ch2 or Ch3
and select psuedocolor for display.
ChD is for transmitted light.
Excitation dichroic
Emission Filters
Detectors
1
0
and 2
0
dichroics
Transmitted light
detector
Meta
Detector
Store
Settings