Reflected-Light Polarization - Detection of Bireflection and Reflection Pleochroism
Application
Polished sections of ores, coals, ceramic products, certain metals and metal alloys display a different
reflection behavior, depending on the orientation of the crystals or the object details. Therefore, this
technique also represents a further contrasting method.
Bireflection
Set the microscope as usual for examinations in reflected-light brightfield (see section 3.2.1)
Pull out plane-glass pushrod (3-3/1).
Insert polarization slider (3-3/3) until stop.
Insert analyzer slider (3-3/4); adjusting lever in center position.
Close aperture diaphragm (3-4/7) by 2/3 of its diameter.
The object displays bireflection if any object details feature differences in brightness and color
which vary when the stage is rotated.
Reflection Pleochroism
Pleochroism can be recognized by color changes in the object after stage rotation (reflected-light
polarizer slider (3-3/3) pushed in, analyzer slider (3-3/4) pulled out). (Rotary λ-compensator pulled out)