Acquiring the image
On the microscope stand - change the light path to
BASEPORT
In Volocity - select channel from pre-set (1)
Refocus / re-centre (microscope focus and joystick)
Then adjust:
2. Exposure time,
3. Binning
4. (and lamp brightness in brightfield channel)
5. NB: Make sure "Auto Contrast" is off!!!
6. Save settings
Repeat for each required channel
Useful tips:
Clicking the "Freeze" button in the software will
keep your actual live image frozen on the
screen; the acquisition stops and the shutters
will close, preventing your sample from
bleaching.
Switch on the contrast range display
(contrast icon on top menu bar).
Left-Click and drag the mouse left
over the image to compress the
displayed range (and right to
expand it). This does not change
the raw data but just the way the
image is displayed on screen