
OPERATION
LSM 510 Image Optimization
B 40-051 e 07/98 4-91
Automatic pinhole adjustment and collimator
The position of the pinhole (X-Y-Z-coordinates) in
relation to the detector makes a major
contribution to image optimization.
In all of the available standard configurations, the
pinhole positions are factory-adjusted and are
activated when a standard configuration is loaded.
The automatic adjustment allows the LSM 510
pinholes to be used with any combination of beam
splitters.
Click on the < Adjust Automatically > button.
The Requirements for Adjustment dialog
box will then appear.
Meet the requirements listed in the dialog box
and press the <OK> button
Pinhole adjustment will then run
automatically. The adjusting procedure takes
approx. 3 min.
The determined data are stored
automatically and will be available for all
further examinations using the same
configuration
Fig. 4-100
Komentarze do niniejszej Instrukcji