OPERATION
LSM 510 Quick start
B 40-051 e 07/98 4-21
(8) Click on the <Stop> button.
(9) Click on the <Optimal Interval> button in the "Optical Slice" window.
The optimum parameters are transferred to the Z-Sectioning section.
(10) Click on the <Start> button. The system will automatically start Z sectioning. Be careful not to
bump the air table or the microscope until z sectioning is completed. Each successive z-slice can be
viewed by changing to the Gallery Mode. This button is located on the right side of the image.
(11) A black bar will be shown under the image and will move from left to right, showing that the
LSM 510 is in the process of Z sectioning. The laser will automatically stop scanning when z-
sectioning is completed.
(12) The entire stack of images can be saved using the <Save> or <Save As> buttons on the right side
of the image.