OPERATION
Annex LSM 510
4-186 B 40-051 e 07/98
Trigger pulse description:
Level detection:
Low level: 0.0 - 1.0 V
High level: 3.0 - 5.0 V
Slew rate:
rising edge: 1 µs
falling edge: 1 µs
Pulse width (always positive pulses/high level):
Trigger-In: 20 ms (Speed 10 - 5)
31 ms (Speed 4)
62 ms (Speed 3)
123 ms (Speed 2)
246 ms (Speed 1)
Trigger-Out: ca. 100 ms
Pulse frequency:
Trigger-In: 2x pulse width
Trigger-Out: > pulse width
Valid edge:
Trigger-In: Rising edge
Trigger-Out: Falling edge
Caution:
Never apply more than 5 V or negative voltages to avoid any damage.
In and outputs are not galvanically decoupled.
Therefore proper measures for galvanic decoupling external devices have to be taken (opto-coupler etc.).
Do not connect no pins declared with ‘reserved’ (see table below). Otherwise at least the interface
can be damaged.