Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
B 40Ć012 e 08/99 IX
The Axiovert 25 microscopes and their original accessories may be used only for the microscopy
techniques described in this manual.
The manufacturer cannot take liability for any other application, including that of individual
modules or components. This also applies to all servicing and repair work which has not been
performed by authorized service personnel. Furthermore, this forfeits all warranty claims also for
those parts which were not directly covered by the repair.
Special note should be taken of the following warnings:
The Axiovert 25 microscopes have been designed, produced and tested in compliance with DIN
EN 61010Ć1 (IEC 1010Ć1), "Safety regulations for electrical measuring, control and laboratory
instruments", including the relevant CSA and UL directives, and left the factory in a perfect
state with regard to its safety facilities.
This operating manual includes information and warnings which must be observed by the user.
The Axiovert 25 microscopes have been designed, in compliance with the latest scientific and
technical development, for the visual, photomicrographic and videotechnological
examination of specimens. The instruments may only be used for their intended application
and have not been designed for unattended constant operation!
The microscopes are not equipped with any special devices for protection from substances
which are corrosive, toxic, radioactive or otherwise hazardous to health.
The wide range power unit which is integrated in the stand of the microscope permits the use
of line voltages in the range between 100 and 240 V AC ±10 %, 50 Ć 60 Hz, without the need
for the voltage to be changed at the instrument.
The Axiovert 25 microscopes are categorized as Protection Class I Equipment.
The power plug must be inserted in a socket featuring a grounding (earth) contact. The
grounding effect must not be nullified by an extension cable which does not have a protective
ground wire. If a transformer is used to adapt the line voltage, it may not nullify the effect of
the ground wire.
Any nullification of the ground wire inside or outside the instrument, or the separation of the
earth contact, may cause danger for the user and is therefore not permitted.