Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
2Ć4
(5) Attachment of the specimen guide M
• Attach the specimen guide M for Axiovert 25
(2Ć6ā/ā6) to the left or right of the stage (2Ć6ā/ā2)
and secure it from below with three knurled
screws (2Ć6ā/ā3).
NOTE The L rail (2Ć6ā/ā4) of the specimen
guide must be flush with the
stage.
The specimen guide is moved by
means of a coaxial drive (2Ć6ā/ā5)
(bottom drive knob for X
movement and top drive knob for
Y movement).
• Select the suitable holding frames (2Ć6ā/ā1) for
the intended specimen vessel and fit the
appropriate scales in the recesses on the
specimen guide and adhere them.
NOTE When using the holding frames, make sure that the holding frames are inserted
correctly in the specimen guide, i.e. the holding frames must engage in the two
holding clips (2Ć6ā/ā7).
• Figure 2Ć24 shows a selection of available holding frames including the appropraite specimen
vessels. The suitable vessels are not, however, included in the delivery package and must be
provided by the customer (see Section 2.5, "Working with the specimen guide M and holding
frames").
Figure 2Ć6 Attachment of the specimen guide M
1
3
4
5
6
7
2