Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
3Ć6
• Use the SW 3 screwdriver to loosen the clamping screw on the heatsink and remove the used
lamp.
The cable on the heatsink must not be removed.
• Hold the new lamp on the marked metal base and insert it in the heatsink in such a way that its
reflecting part is at the bottom when inserted in the mount. The reflecting meltĆoff part of the
lamp must point to the side (to avoid interference with the image).
• Carefully tighten the clamping screw on the heatsink.
• Avoid fingerprints on the glass parts of the lamp; if required, remove fingerprints immediately.
• Press down spring lever and insert the lamp in the mount while holding it on the heatsink.
• Slowly release pressure from spring lever and let go of the heatsink.
The heatsink must be aligned parallel to the lamp housing. The position of the
heatsink can be corrected by pressing down the spring lever and turning the heatsink.
• Attach the lamp housing to the lamp base and tighten screw (3Ć4ā/ā1).
• Attach the microscope illuminator to the microscope stand.
• Note down the number indicated on the operating hours counter of the power supply unit. The
HBO lamp must be replaced when the nominal life of 100 hours has been reached.