Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
B 40Ć012 e 08/99 2Ć13
NOTE Should the setting range of the centering screws not be sufficient, please check
whether the condenser is at the front stop and the Ph/H/Ph slider has engaged. Exact
imaging of the annular diaphragm requires a planeĆparallel object.
• Replace the auxiliary lens with the eyepiece
again.
• The center position of the Ph/H/Ph slider
permits brightfield illumination, the brightĆ
ness of which is matched to that of the phaseĆ
contrast image via an attenuation filter (gray
filter) (2Ć17ā/ā2).
• When using different Ph objectives with Ph0,
Ph1 and Ph2, pay attention to the correct
allocation of the annular diaphragms (see
table).
Objective Ring and VAREL diaphragms for
Condenser 0.55 Condenser 0.4 Condenser 0.2
CPĆAchromat 5xā/ā0.12 Ph0 -- Ph0 Ć 0.4 451255 Ph0 Ć 0.2 451261
CPĆAchromat 10xā/ā0.25 Ph1 Ph1 Ć 0.55 451258 Ph1 Ć 0.4 451256 Ph1 Ć 0.2 451260
CPĆAchromat 10xā/ā0.25 Ph1ā/āVar1 Ph1 Ć 0.55 451258
Var1 Ć 0.55 451252
Ph1 Ć 0.4 451256
Var1 Ć 0.4 451251
LDĆAchrostigmat 20xā/ā0.30 Ph1 Ph1 Ć 0.55 451258 Ph1 Ć 0.4 451256 Ph1 Ć 0.2 451260
LDĆAchrostigmat 20xā/ā0.30 Ph1ā/āVar1 Ph1 Ć 0.55 451258
Var1 Ć 0.55 451252
Ph1 Ć 0.4 451256
Var1 Ć 0.4 451251
LDĆAchrostigmat 32xā/ā0.40 Ph1 Ph1 Ć 0.55 451258 Ph1 Ć 0.4 451256 Ph1 Ć 0.2 451260
LDĆAchrostigmat 32xā/ā0.35 Ph1ā/āVar1 Ph1 Ć 0.55 451258
Var1 Ć 0.55 451252
Ph1 Ć 0.4 451256
Var1 Ć 0.4 451251
LDĆAchrostigmat 40xā/ā0.55 Ph2 Ph2 Ć 0.55 451259 Ph2 Ć 0.4 451257
LDĆAchrostigmat 40xā/ā0.50 Ph2 Var2 Ph2 Ć 0.55 451259
Var2 Ć 0.55 451253
• Loosen the centering screws to allow the inserted annular diaphragm to be exchanged for
another one.
Figure 2Ć17 Insertion of the ring diaphragm
1
2
3
3