Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
B 40Ć012 e 08/99 3Ć5
Safety notes for use
CAUTION The HBO 50 is under high pressure during operation. It may therefore only be
operated in a closed microscope illuminator.
• The cooling process of the lamp housing must not be obstructed by covers.
• The HBO 50 must cool down for approx. 15 minutes before the lamp can be changed.
• The lamp emits UV radiation during operation. Protect your eyes and skin from this radiation.
It is recommended to wear safety goggles when handling the microscope illuminator.
CAUTION The HBO 50 must be replaced after expiry of its average life of 100 h. When the
average life is exceeded, the risk of explosion increases.
The operating time of the HBO 50 can be read from the hour meter.
This warning plate on the rear of the illuminator means:
Caution: hot surface!
Allow the lamp to cool down before touching it.
Change of lamp
CAUTION Switch off power unit and remove
plug of lamp mount from the
socket of the power unit. Allow
lamp and lamp housing to cool
down (approx. 15 min.).
• Remove HBO 50 illuminator from the
microscope stand after loosening the
clamping screw.
• Place the illuminator on a flat work surface.
• Loosen screw (3Ć4/1) using the SW 3 screwĆ
driver supplied and remove the lamp housing
(3Ć4/2).
• Press down spring lever (3Ć4/5), hold lamp
(3Ć4/4) on heatsink (3Ć4/3) and pull it out of
the lamp mount.
• Place the heatsink with lamp on the work
surface in such a way that the clamping screw
on the heatsink is accessible.
Figure 3Ć4 Change of HBO 50 lamp
1
2
3
4
6
5