Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
2Ć22
2.8.1 Glass stage
The glass stage (2Ć28ā/ā3) enables convenient
working with simultaneous observation of the
objectives and easily meets the high demands
on cleanliness.
Installation
• Fit the glass stage onto the stage coupling
point using the spacers.
• Secure the glass stage on the stand using the
securing screws (2Ć28ā/ā2).
• Seal the rear holes on the surface of the glass
stage by means of the cover (2Ć28ā/ā1).
NOTE When using the glass stage, make sure that the objective front is no more than max.
3 mm above the stage surface in order to avoid hitting the stage when the objectives
are changed. If necessary, the nosepiece must be lowered.
2.8.2 Heating stage
A heating stage is available for the Axiovert 25 microscope. It consists of:
• Heating microscope stage M (471820 9901), three spacers are contained in the package
• Hot plate (471841) and
• TempControl 37Ć2 (471842).
Asssembly must be performed in accordance with section 2.8.
For the operation of the heating microscope stage, please see manual G 42Ć527.
The air temperature unit (471824) can also be used to control the temperature of the object spot
in the imaging beam path.
Figure 2Ć28 Insertion of the glass stage
1
2
3