Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
B 40Ć012 e 08/99 3Ć3
(2) Change of halogen lamp
Proceed as follows to change the lamp:
Disconnect the instrument from the line.
Unlock the housing (3Ć2/1) by slightly turning it counterclockwise and remove it.
Remove the halogen lamp with adjustment base (3Ć2/2) after a cooling time of approx.
15Ăminutes.
Remove new 6 V, 30 W halogen lamp with adjustment base from the packing box and insert it
in the mount; the tip of the mount must engage in the centering notch of the carrier plate.
IMPORTANT Do not touch the lamp bulb with your bare hands; if required, clean the bulb using
pure alcohol before switching on for the first time in order to stop dust from burning
in.
Attach housing again after changing the halogen lamp and lock it.
Figure 3Ć2 Change of halogen lamp