3 Maintenance and troubleshooting
3.1 Maintenance
Maintenance of the the Axiovert 25 microscope is limited to the following operations:
Cover the instrument with the dust cover after every use.
Clean exposed optical components whenever required.
Carefully remove moisture deposit or precipitated aggressive vapor using a dry cloth.
Protect the instrument from temperatures above 50 °C, frost, humidity, chemically aggressive
vapor/substances.
Remove dust from optical surfaces using a rubber blower or a natural hair brush which can be
degreased in alcohol and dried afterwards. Remove stubborn dirt and fingerprints using a dustĆ
free cloth or leather; breathe on the dirty surface, if required.
Clean the front lenses of the objectives using petroleum ether, but do not use alcohol.
Use commercially available optics cleaning cloths to remove stubborn dirt; if necessary, lightly
moisten the cloths with petroleum ether.
When using the Axiovert 25 microscope in humid climatic zones, proceed as follows:
Store the Axiovert 25 microscope in bright, dry and well ventilated rooms with a humidity of less
than 85 %; store particularly sensitive components and accessories, such as objectives and
eyepieces, in a dry closet.
When storing the equipment in closed cases for a longer period of time, the growth of fungus
can be avoided by including cloths soaked in fungicide in the cases.
NOTE The risk of growth of fungus on optoĆmechanical instruments always exists in the follĆ
lowing conditions:
relative humidity of more than 75 % and temperatures between +15
°
C and +35ðC
for more than 3 days;
installation in dark rooms without air ventilation;
dust deposits and fingerprints on optical surfaces.