Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
B 40Ć012 e 08/99 2Ć9
2.3 Use of LD objectives
When inverted microscopes are used for analyses, the bottom thickness of the vessels used differs
considerably from the standard cover slip thickness of 0.17 mm.
Normally, such distances are easily bridged by the working distances of lowĆpower objectives:
- CP Achromat 5x/0.12 FWD (in air) 10.8 mm or
- CP Achromat 10x/0.25 FWD (in air) 5.1 mm.
Already in the middle magnification range, however, these working distances are reduced to values
around or below 1 mm. Such objectives can then no longer be used for thicker bottoms.
This shortcoming is remedied by special LD
(Long Distance) objectives; these objectives
feature a relatively long working distance, but
also the normal 45 mm parfocalization length of
all other objectives.
For adaptation to the bottom thickness in use,
fit the corresponding cover glass cap (2Ć12ā/ā1)
onto the objective (see table below for assignĆ
ments).
Table of free working distances on the Axiovert 25
Objective name Order No. FWD
without cover glass cap
for bottom thickness
FWD
with cover glass cap
for bottom thickness
D=0.17
[mm]
D=1
[mm]
D=1.2 - 1.8
[mm]
0 Ć 0.6
(444931)
[mm]
0.6 Ć 1.2
(444930)
[mm]
CPĆAchromat 5xā/ā0.12 Ph0 440921 10.7 10.3 10.2 ... 9.9* - -
CPĆAchromat 10xā/ā0.25 Ph1 440931 5.0 4.6 4.5 ... 4.2* - -
CPĆAchromat 10xā/ā0.25 Ph1/Var1 440938 5.0 4.6 4.5 ... 4.2* - -
LDĆAchrostigmat 20xā/ā0.30 Ph1 440147 2.2 1.8 1.7 ... 1.4* - -
LDĆAchrostigmat 20xā/ā0.30 Ph1/Var1 440148 2.2 1.8 1.7 ... 1.4* - -
LDĆAchrostigmat 32xā/ā0.40 Ph1 440149 2.5 2.1 2.0 ... 1.7* - -
LDĆAchrostigmat 32xā/ā0.40 Ph1/Var1 440146 2.5 2.1 2.0 ... 1.7* - -
LDĆAchrostigmat 40xā/ā0.55 Ph2 440155 - - 1.8 ... 1.5* 1.6 ... 1.3* 1.6 ... 1.3*
* depending on bottom thickness
The working distances of PhVar objectives correspond to those of the Ph objectives.
Figure 2Ć12 Working with LD objectives